Please Choose Your Language:

English
Español

Thermostatic Incubator

Thermostatic Incubator

The thermostatic incubator is an essential experimental equipment in biology, genetic engineering, medicine, health and epidemic prevention, environmental protection, agriculture, forestry, animal husbandry, and other industries. It is widely used in constant temperature experiments, culture experiments, ecological testing, etc.

Details of the thermostatic incubator:

Thermostatic Incubator details

Product Features:
1. Inner mirror 304 SS material, corrosion resistance.
2. Inner fan installed, force the air circulation, so good temperature uniformity.
3. Good quality 304 stainless steel shelf, good load capacity.
4. Double door design, it is easy to observe the culture.
5. Intelligent LCD control panel, excellent temperature control performance, and easy operation.

Temperature controller details:

Thermostatic Incubator details

1. Adopt PID control, low-temperature fluctuation, and a programmable controller. The user can preset 15 segments, and each program has 30 steps. Each timing segment is from 1- 99 hours and 59 minutes.
2. The machine will automatically send alarms by sound and light if over temperature.

Other Details:

Thermostatic Incubator details

Technical Data of the Thermostatic Incubator:

Temperature control range: room temperature+5℃-65℃
Temperature control resolution:0.1℃
Temperature control fluctuation:0.3℃
Timing:1-9999 min
Shelves:2 pcs
Power Supply:220V 50 Hz

ModelMHP - 9052MHP - 9082MHP - 9162MHP - 9272
Input Power (W)250350550700
Volume (L)5080160270
Working Chamber Size (mm)345 *355 *410400 *400 *500500 *500 *650600 *600 *750
Overall Size (mm)475 *460 *695530 *505 *785630 *605 *935730 *705 *1035

How to use the thermostatic incubator:

1. Power on the thermostatic incubator.
2. Temperature setting: press the set key to set the temperature or look up for the set temperature. Press the set key once, and the digital shines, meaning it has entered the setting mode. Press the up key to add the value and the down key to subtract the value. Long press the key to add or subtract the value continuously. Press the set key again to go back to normal mode.

Precaution of using the illumination incubator:

1. The incubator should be adequately grounded.
2. The incubator should be put on stable ground if in use and put in cool, dry, and ventilated places after use. Try to avoid direct sunlight.
3. To ensure effective heat dissipation of the condenser, the incubator should be kept at least 10 cm from the wall; if possible, 15 cm is an ideal distance. The side wall of the incubator should be kept at a distance of at least 50mm, and 300mm from the top wall.
4. Try to handle carefully during transportation and try to avoid shock. The inclined angle should be less than 45 degrees.
5. Please check the fuse if the instrument does not work.